When designing an XPS experiment, one of the first things you must think about is which X-ray source you need. Most commonly, monochromated aluminium kα radiation (1486.7 eV) is used as the exciting source however magnesium kα (1253.6) has been used extensively also. While magnesium sources are non-monochromatic, the intrinsic line width of Mg kα is lower than Al kα (0.7 vs 0.85 eV) and hence is the preferred choice of non-monochromatic sources.

Most modern XPS instruments will come with monochromated Al kα as standard, but may also feature a non-monochromated dual anode source (typically Mg/Al) or even a monochromated Ag Lα/Al kα source (such as the Kratos Supra at HarwellXPS).

Monochromatic vs non-monochromatic X-rays

Monochromatic X-rays are typically preferable since the narrow X-ray line width results in enhanced resolution of spectral features and thus are capable of resolving chemical states with smaller differences in binding energy. The drawback of monochromatic X-rays is of course reduced signal, since you are removing photons from the exciting source, however this is a non-issue with monochromatic Al kα sources, where focusing crystals or focused beams often result in comparable or even high signal than unfocused, non-monochromatic X-rays.

Sampling depth

The different energies associated with different X-ray sources result in different kinetic energies and hence different inelastic mean free path lengths. When considering the energies involved with XPS you can generally assume that higher energies result in deeper sampling depths, however this might not hold true for certain orbitals which result in very low kinetic energies.

The universal curve for IMFP is given by λ = 143/E2 + 0.054√ E (Figure 1).(1)

Figure 1: Compilation of IMFP of elements in nanometres by Seah and Dench.(1)

One major advantage of this variation in IMFP can be found when analysing thin films or similar, in which simultaneous equations of intensity attenuation for each source and path length may be used to calculate film thickness.(2,3)

The top end of lab based systems in terms of depth is the liquid Ga source (see the HarwellXPS associated facility) which may reach depths of up to 30-60 nm.

Auger overlap

It is not uncommon in XPS to find photoemission peaks of interest overlapping with auger lines from other elements present in the sample (or even the same element in the case of cobalt/Al kα). In this case, analysis using another X-ray energy can help remove this overlap.

This is a possibility since the kinetic energy of an auger electron is independent of exciting energy. Therefore, when you increase the energy of the incoming radiation, you increase the kinetic energy of the photoelectrons (equation 1) and thus separate auger peak from photoemission.

hv = EK + EB + Φ Equation 1

X-ray source energies

Commonly used X-ray sources are listed in table 1.

AnodeRadiationEnergy / eVLine width / eV
MgKα1253.60.7
AlKα1486.70.85
ZrLα2042.41.6
AgLα2984.32.6
TiKα4510.92.0
CrKα54172.1
GaKα9250
Table 1: Common X-ray source energies(4)

References

  1. Seah, M. P. and W. Dench (1979). “Quantitative electron spectroscopy of surfaces: A standard data base for electron inelastic mean free paths in solids.” Surface and interface analysis 1(1): 2-11. Read it online here.
  2. Isaacs, M. A., et al. (2017). “Tunable Ag@ SiO 2 core–shell nanocomposites for broad spectrum antibacterial applications.” RSC advances 7(38): 23342-23347. Read it online here.
  3. Shard, A. G. (2012). “A straightforward method for interpreting XPS data from core–shell nanoparticles.” The Journal of Physical Chemistry C 116(31): 16806-16813. Read it online here.
  4. XPS Simplified, accessed 14:05 29/05/2020

Was this article helpful?

Yes No

Related Articles

  • Pass Energy
    • 4
    • 336
  • Argon Clusters vs Monatomic Ions
    • 0
    • 288
  • XPS Imaging
    • 1
    • 399
  • Analysis Depth in XPS & IMFP
    • 0
    • 373
  • Detection Limits in XPS
    • 0
    • 332
  • Experimental Reporting
    • 0
    • 292

Leave A Comment? Cancel Reply

User Login

Log In
Register Lost Password

Forums

  • General Discussion Forum
  • XPS Analysis Software

Post Categories

  • General
  • Knowledge
  • Review
  • Uncategorized

Forum Statistics

Registered Users
38
Forums
6
Topics
7
Replies
5
Topic Tags
5

HarwellXPS Guru

(c) 2020 HarwellXPS

Follow Us

Copyright HelpGuru. Powered by HeroThemes.